Metrology for Drug Delivery project – results and impact

Elsa Batista, Hugo Bissig, John Morgan, Anders Niemann, Annemoon Timmerman, Florestan Ogheard
Abstract:
This paper presents the major impacts of the work developed in EMPIR Project - MeDD II, Metrology for Drug Delivery in the reference standards used for flow rate determination of drug delivery devices and the results obtained in each work package, with a special focus on the primary standards developed by the partners using different technologies. These new primary standards were validated via an inter-laboratory comparison using two different type of flow meters and a precision syringe pump, in a range from 1500 nL/min down to 5 nL/min. This inter-comparison was performed between nine participating laboratories, each with different methodologies, measurement principles, flow ranges,operating conditions, and measurement uncertainty values.
Keywords:
Drug delivery; New primary standard; Inter-laboratory comparison
Download:
IMEKO-TC9-2019-146.pdf
DOI:
10.21014/tc9-2022.146
Event details
IMEKO TC:
TC9
Event name:
FLOMEKO 2022
Title:

19th International Flow Measurement Conference 2022

Place:
Chongqing, CHINA
Time:
01 November 2022 - 04 November 2022