Evaluation of Microflow Calibration Rig using Static Weighing System with Flying Start-and-Finish Method |
---|
R. Doihara, K.-H. Cheong, N. Furuichi |
- Abstract:
- A Liquid microflow calibration rig was designed and evaluated. The developed calibration rig has a static weighing system and improved diverter valves for small flow rates. The weighing system succeeded in reducing an uncertainty due to an evaporation effect by using a detachable weighing tank. Experimental results of evaporation effect were introduced. In this study, two types of devices were introduced so that a Flying Start-and-Finish (FSF) method was implemented into the static weighing system. One of these devices is a set of diverting valves, and the other is a syringe pump with a linear encoder. Estimation tests were conducted. Diverter timing errors for the set of diverting valves were estimated to be about 6 ms.
- Keywords:
- Micro-flow; Liquid flow; Calibration; Diverter
- Download:
- IMEKO-TC9-2019-141.pdf
- DOI:
- 10.21014/tc9-2022.141
- Event details
- IMEKO TC:
- TC9
- Event name:
- FLOMEKO 2022
- Title:
19th International Flow Measurement Conference 2022
- Place:
- Chongqing, CHINA
- Time:
- 01 November 2022 - 04 November 2022