SIGNIFICANCE OF ORTHOGONALITY IN MASS CALIBRATION |
---|
Woo Gab Lee, Jin Wan Chung, Kwang Pyo Kim |
- Abstract:
- The designs for submultiples of the unit of mass have been studied to find designs where two comparators are used. This study has found several cases where orthogonal designs can be used for using two comparators in a decade. Even though the orthogonal designs are for the case where the ratio of the comparator’s standard deviation is 2, simulation showed extended values of the ratio could be used to reduce variances with the orthogonal designs.
- Keywords:
- calibration design, orthogonality
- Download:
- PWC-2003-TC3-017.pdf
- DOI:
- -
- Event details
- Event name:
- XVII IMEKO World Congress
- Title:
Metrology in the 3rd Millennium
- Place:
- Dubrovnik, CROATIA
- Time:
- 22 June 2003 - 28 June 2003