EXPERIMENTAL VERIFICATION OF CALCULATED FREQUENCY CHARACTERISTICS OF OCTOFILAR RESISTORS |
---|
Jaroslav Bohácek |
- Abstract:
- With an intention to reveal possible inaccuracies of mathematical models used in calculations of frequency characteristics of resistors of octofilar design, comparisons of a 12 906 Ω and a 1 Ω octofilar resistor against monofilar and quadrifilar calculable resistors of the same nominal values have been performed. A possibility to carry out these precision comparisons by means of a commercially available LCR meter has been investigated.
- Keywords:
- calculable resistor, octofilar design, LCR meter
- Download:
- PWC-2003-TC8-001.pdf
- DOI:
- -
- Event details
- Event name:
- XVII IMEKO World Congress
- Title:
Metrology in the 3rd Millennium
- Place:
- Dubrovnik, CROATIA
- Time:
- 22 June 2003 - 28 June 2003