Analysis of flow rate measurement accuracy and traceability of flowmeters in field conditions using clamp-on ultrasonic flowmeters |
---|
C. Simões, A. Ribeiro, M. Almeida, D. Covas |
- Abstract:
- This study examines the traceability and measurement uncertainty of in situ hydraulic calibration using clamp-on ultrasonic flowmeters as a reference. The procedure compares the equipment readings with the reference ones. Measurement uncertainty evaluation uses GUM formulation, considering the linearity conditions of the mathematical models applied. Experimental values are used to test the procedure and its suitability for actual cases where the expected accuracy needs to be achieved.
- Keywords:
- Flow measurement, clamp-on ultrasonic flowmeter, measurement uncertainty, traceability, measurement accuracy.
- Download:
- IMEKO-TC11-2023-04.pdf
- DOI:
- 10.21014/tc11-2023.04
- Event details
- IMEKO TC:
- TC11
- Event name:
- IMEKO TC8, TC11 and TC24 Conference
- Title:
Testing, Inspection and Certification for Confidence and Safety
- Place:
- Madeira, PORTUGAL
- Time:
- 11 October 2023 - 13 October 2023