Novel quality assessment protocol based on Kiviat diagram for pulsed wave Doppler diagnostic systems: first results |
---|
Giorgia Fiori, Gabriele Bocchetta, Maurizio Schmid, Silvia Conforto, Salvatore Andrea Sciuto, Andrea Scorza |
- Abstract:
- Ultrasound (US) systems are routinely and extensively used in several medical fields despite the lack of an internationally accepted quality standard for their testing. One crucial aspect for Quality Assessment (QA) protocols is the need to summarize the contribution of the large number of existing test parameters in a few meaningful quantities. The study herein proposed fits into this context, focusing on the use of the Kiviat diagram applied to Pulsed Wave Doppler (PWD) equipment QA. Four test parameters, derived from the literature, were objectively assessed through custom-written image analysis-based methods and then scaled for an effective combination. The experimental setup used to collect PWD spectrograms included an intermediate technology level US diagnostic system equipped with a linear array probe and a commercial Doppler flow phantom. Tests were repeated at two Doppler frequencies.
- Keywords:
- quality assessment protocol, Kiviat diagram, pulsed wave, Doppler diagnostic systems
- Download:
- IMEKO-TC4-2023-38.pdf
- DOI:
- 10.21014/tc4-2023.38
- Event details
- IMEKO TC:
- TC4
- Event name:
- TC4 Symposium 2023
- Title:
26th IMEKO TC4 Symposium and 24th International Workshop on ADC and DAC Modelling and Testing (IWADC)
- Place:
- Pordenone, ITALY
- Time:
- 20 September 2023 - 21 September 2023