Evaluation results of the new standard gas flow system at VMI: Piston prover system

Nguyen Ngoc Hai, Nguyen Xuan Thai, Phan Lac Tuan, Duong Hong Son
Abstract:
Currently many National Metrology Institute (NMIs) as well as advanced calibration laboratories are using piston gas flow standards with mercury sealing method for gas flow calibration in the low pressure range. The flow in the small range is about cc/min. In this work, the low pressure gas flow calibration system at VMI is presented, designed and manufactured by the Taiwan National Metrology Institute, CMS/ITRI. The flow range is within (0.002 - 24) L/min. The uncertainty of the reference system is assessed against the ISO/IEC Guide 98-3:2008 document, Uncertainty is evaluated from individual influence sources such as category A and B assessments. The standard uncertainty/relative standard uncertainty and degrees of freedom of the sources can be evaluated individually and then combined to produce a composite standard uncertainty/combined relative standard uncertainty and an effective degree of freedom. Finally, the relative expanded uncertainty is obtained by multiplying the relative standard uncertainty associated with a coverage factor at the 95 % confidence level of the measurement result.
Keywords:
Air speed; Gas flow; LDA; Piston
Download:
IMEKO-TC9-2019-055.pdf
DOI:
10.21014/tc9-2022.055
Event details
IMEKO TC:
TC9
Event name:
FLOMEKO 2022
Title:

19th International Flow Measurement Conference 2022

Place:
Chongqing, CHINA
Time:
01 November 2022 - 04 November 2022