High capacitance simulation using mutual inductors
Stanislav Mašlán
Abstract:
This paper describes experimental high capacitance simulator based on a mutual inductance. The standard was designed for calibration of battery impedance analyzers (EIS). The designed standard is equipped by a DC bias voltage source designed to simulate typical lithium battery cell. The DC bias source was designed so it does not affect properties of simulated impedance by its own internal impedance. The whole standard was designed for a frequency range from fraction of hertz to at least 5 kHz and it can withstand peak measurement current up to 2.5 A.