THE COMPARISON OF THE REALISATION OF THE FREEZING POINT OF INDIUM INSIDE THE THREE ZONE FURNACE AND THE FLUDISED POWDER FURNACE |
---|
Jovan Bojkovski, Valentin Batagelj, Igor Pušnik |
- Abstract:
- In the paper the comparison of the realisation of the freezing point of indium inside the three-zone furnace and the fluidised powder furnace is presented. The emphasis of the paper is given to the comparison of the metrological characteristics of the realisations such as time duration of the plateau and the width of the plateau. The same fixed-point cell (closed fixed-point cell, serial number In 86, made by Isotech) was realised in both furnaces. Also the analysis of the immersion curve is performed in order to see how it agrees with the expected curve.
- Keywords:
- temperature, freezing point of indium, comparison of realisations
- Download:
- PWC-2003-TC12-024.pdf
- DOI:
- -
- Event details
- Event name:
- XVII IMEKO World Congress
- Title:
Metrology in the 3rd Millennium
- Place:
- Dubrovnik, CROATIA
- Time:
- 22 June 2003 - 28 June 2003