APPLICATION OF SELF-CALIBRATING THERMOCOUPLES WITH MINIATURE FIXED-POINT CELLS IN A TEMPERATURE RANGE FROM 500°C TO 650°C IN STEAM GENERATORS |
---|
Frank Bernhard, Dirk Boguhn, Silke Augustin, Helge Mammen, Andrej Donin |
- Abstract:
- Using integrated miniature fixed-point cells, a measuring uncertainty of < 1 K can be reached under operating condi tions in the superheated steam range of power plants by a periodic recalibration of the thermocouples, with operating times of > 20000 h. The fixed-point materials used for a temperature range from 500°C to 650°C are technically pure metals and binary alloys.
- Keywords:
- temperature measurement, measuring uncertainty, self-calibration, fixed-point, superheated steam
- Download:
- PWC-2003-TC12-005.pdf
- DOI:
- -
- Event details
- Event name:
- XVII IMEKO World Congress
- Title:
Metrology in the 3rd Millennium
- Place:
- Dubrovnik, CROATIA
- Time:
- 22 June 2003 - 28 June 2003