INFLUENCE OF EXCITATION VOLTAGE OVER METROLOGICAL PERFORMNACE OF FORCE TRANSDUCERS |
|---|
| Harish Kumar, Anil Kumar |
- Abstract:
- The present study discusses the effect of excitation voltage of the digital indicator over the metrological performance of the force transducers. The force transducers of 1 kN, 2 kN and 5 kN nominal capacity and having relative repeatability error better than ± 0.002 % have been calibrated on the basis of calibration procedure based on ISO 376-2004. The relative deviations due to repeatability and reproducibility error have been computed for the force transducers and results have been reported.
- Keywords:
- force transducer, excitation voltage, ISO 376-2004, metrological performance
- Download:
- IMEKO-WC-2012-TC3-P11.pdf
- DOI:
- -
- Event details
- Event name:
- XX IMEKO World Congress
- Title:
Metrology for Green Growth
- Place:
- Busan, REPUBLIC of KOREA
- Time:
- 09 September 2012 - 12 September 2012