MICROMECHANICAL SILICON MICROBALANCE

K. Riski, A. Oja, T. Sillanpää, H. Sipola, H. Seppä, J. Kiihamäki, P. Seppälä, J. Karttunen
Abstract:
A micromachined capacitive silicon microbalance has been designed and fabricated. It is intended for weighing masses of the order of 1 g with a resolution and accuracy of about 1 µg. The device consists of a micromachined SOI (silicon-on-insulator) chip which is anodically bonded to a glass plate. The capacitance is formed between two electrodes. The flexible electrode is the SOI layer. The other electrode is metal layer on the glass. The glass electrode is divided into three sections. The sections are used for detecting the tilting of the top electrode due to eccentric loading. The measuring circuit implements electrostatic force feedback which keeps the top electrode at a constant horizontal position irrespective of the mass of the load. First test measurements have demonstrated that an accuracy of about 2 to 3 µg at 1 g can be reached.
Download:
IMEKO-TC3-2001-031.pdf
DOI:
-
Event details
IMEKO TC:
TC3
Event name:
TC3 Conference 2001
Title:

17th Conference on Force, Mass and Torque

Place:
Istanbul, TURKEY
Time:
17 September 2001 - 21 September 2001