A METHOD OF TRACEABILITY FOR A FPG8601 FORCE BALANCED PISTON GAUGE TO DEFINE PRESSURES IN THE RANGE FROM 1 PA TO 15 KPA IN GAUGE AND ABSOLUTE MEASUREMENT MODES |
|---|
| Rob Haines, Michael Bair |
- Abstract:
- The FPG8601 has become an important reference for many primary measurement laboratories in a pressure range from approximately 1 Pa to 15 kPa in both gauge and absolute measurement modes. The FPG8601 measurement range spans the gap between the traditional measurement regimes of "pressure" and "'vacuum", a range where other primary pressure standards either do not exist or do not have uncertainties sufficient to support the FPG8601 traceability. The limitations of available primary standards create a challenge for FPG8601 users needing to define traceability in this range. This paper suggests a method of maintaining traceability using force, dimensional and primary pressure measurements in a manner different from that normally used for traditional "floating" piston gauges.
- Keywords:
- pressure, vacuum, traceability
- Download:
- IMEKO-WC-2009-TC16-298.pdf
- DOI:
- -
- Event details
- Event name:
- XIX IMEKO World Congress
- Title:
Fundamental and Applied Metrology
- Place:
- Lisbon, PORTUGAL
- Time:
- 06 September 2009 - 11 September 2009