Measurement of gravitational acceleration values at the calibration laboratories in Korea

S.Y.Woo, I.M.Choi, H.W.Song
Abstract:
The most important conversion factor between mass and force is the value g of the acceleration due to gravity. This can be measured with an absolute or relative gravimeter. It depends on the latitude and height above sea level. Values of g at many calibration laboratories in Korea are measured with a relative gravimeter, Model G manufactured by La Coste & Romberg in USA. To improve the measurement performance of the gravimeter, we attached an optical magnifying system instead of microscope eyepiece, which consists of CCD camera and LCD display. This optical system showed to be sufficiently convenient and accurate to provide a practical alternative to the traditional microscopic eyepiece. The observed value of g spans a range of 0.03% between the northernmost and the southernmost calibration laboratory in Korea. In this paper, the improved performance and the detailed gravity measurement results are given.
Keywords:
gravity, acceleration, gravimeter, calibration
Download:
IMEKO-TC3-2007-112u.pdf
DOI:
-
Event details
IMEKO TC:
TC3
Event name:
Cultivating Metrological Knowledge
Title:

20th Conference on Measurement of Force, Mass and Torque (together with 3rd TC16 Conference on Pressure Measurement & 1st TC22 Conference on Vibration Measurement)

Place:
Merida, MEXICO
Time:
27 November 2007 - 01 December 2007