THE EFFECT OF DIFFERENT INLET-VELOCITY PROFILES ON THE PERFORMANCE OF A CORIOLIS FLOWMETER |
---|
Gregor Bobovnik, Jože Kutin, Ivan Bajsic |
- Abstract:
- Numerical simulations (using finite volume method) of the flow of a viscous fluid through a measuring tube were performed to identify the effects of different inlet-flow conditions on the performance of a straight, slendertube, Coriolis flowmeter. The magnitudes of the antisymmetric fluid forces and the twisting moments acting on the measuring tube were compared with results from a onedimensional fluid-flow model. Simulations were made for some hypothetical inlet-velocity profiles that represent extreme cases of various flowmeters’ installation positions and flow regimes in a pipeline. No significant effects of the modelled inlet-velocity profiles on the performance of the flowmeter were observed.
- Keywords:
- Coriolis flowmeter, velocity profile, numerical simulation
- Download:
- PWC-2003-TC9-005.pdf
- DOI:
- -
- Event details
- Event name:
- XVII IMEKO World Congress
- Title:
Metrology in the 3rd Millennium
- Place:
- Dubrovnik, CROATIA
- Time:
- 22 June 2003 - 28 June 2003